Development and Validation of Endovascular Chemotherapy Filter Device for Removing High-Dose Doxorubicin: Preclinical Study

Author:

Patel Anand S.1,Saeed Maythem2,Yee Erin J.2,Yang Jeffrey2,Lam Gregory J.2,Losey Aaron D.2,Lillaney Prasheel V.2,Thorne Bradford2,Chin Albert K.3,Malik Sheena3,Wilson Mark W.2,Chen Xi C.4,Balsara Nitash P.567,Hetts Steven W.2

Affiliation:

1. Department of Radiology and Biomedical Imaging, University of California San Francisco, 185 Berry Street, Suite 350, San Francisco, CA 94107-5705 e-mail:

2. Department of Radiology and Biomedical Imaging, University of California San Francisco, San Francisco, CA 94107

3. ChemoFilter, Inc., 645 Woodstock Road, Hillsborough, CA 94010

4. Materials Sciences Division, Lawrence Berkeley National Laboratory, Berkeley, CA 94720

5. Materials Sciences Division, Lawrence Berkeley National Laboratory, Berkeley, CA 94720;

6. Environmental Energy Technologies Division, Lawrence Berkeley National Laboratory, Berkeley, CA 94720;

7. Department of Chemical and Biomolecular Engineering, University of California, Berkeley, CA 94720

Abstract

To develop a novel endovascular chemotherapy filter (CF) able to remove excess drug from the blood during intra-arterial chemotherapy delivery (IAC), thus preventing systemic toxicities and thereby enabling higher dose IAC. A flow circuit containing 2.5 mL of ion-exchange resin was constructed. Phosphate-buffered saline (PBS) containing 50 mg doxorubicin (Dox) was placed in the flow model with the hypothesis that doxorubicin would bind rapidly to resin. To simulate IAC, 50 mg of doxorubicin was infused over 10 min into the flow model containing resin. Similar testing was repeated with porcine serum. Doxorubicin concentrations were measured over 60 min and compared to controls (without resin). Single-pass experiments were also performed. Based on these experiments, an 18F CF was constructed with resin in its tip. In a pilot porcine study, the device was deployed under fluoroscopy. A control hepatic doxorubicin IAC model (no CF placed) was developed in another animal. A second CF device was created with a resin membrane and tested in the infrarenal inferior vena cava (IVC) of a swine. In the PBS model, resin bound 76% of doxorubicin in 10 min, and 92% in 30 min (P < 0.001). During IAC simulation, 64% of doxorubicin bound in 10 min and 96% in 60 min (P < 0.001). On average, 51% of doxorubicin concentration was reduced during each pass in single pass studies. In porcine serum, 52% of doxorubicin bound in 10 min, and 80% in 30 min (P < 0.05). CF device placement and administration of IAC were successful in three animals. No clot was present on the resin within the CF following the in vivo study. The infrarenal IVC swine study demonstrated promising results with up to 85% reduction in peak concentration by the CF device. An endovascular CF device was developed and shown feasible in vitro. An in vivo model was established with promising results supporting high-capacity rapid doxorubicin filtration from the blood that can be further evaluated in future studies.

Publisher

ASME International

Subject

Biomedical Engineering,Medicine (miscellaneous)

Cited by 18 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献