Non-Contact Irreversible Electroporation in the Esophagus With a Wet Electrode Approach

Author:

Sheehan Mary Chase1ORCID,Collins Scott2,Wimmer Thomas3,Gutta Narendra Babu4,Monette Sebastian5,Durack Jeremy C.6,Solomon Stephen B.7,Srimathveeravalli Govindarajan8

Affiliation:

1. Department of Mechanical Engineering, Amherst Life Sciences Laboratories, University of Massachusetts , 240 Thatcher Road, Amherst, MA 01003

2. Department of Biomedical Engineering, Amherst Life Sciences Laboratories, University of Massachusetts , 240 Thatcher Road, Amherst, MA 01003

3. Department of Radiology, Division of General Radiology, Medical University of Graz , Auenbruggerplatz 9, Graz 8036, Austria

4. 70 Doctors Park Dr, Cape Girardeau, MO 63703

5. Laboratory of Comparative Pathology, Memorial Sloan Kettering Cancer Center, Weill Cornell Medical College, The Rockefeller University , New York, NY 10065

6. Ajax Health , 330 Madison Avenue, Suite 200, New York, NY 10017

7. Department of Radiology, Interventional Radiology Service, Memorial Sloan Kettering Cancer Center , New York, NY 10065

8. Department of Mechanical Engineering Institute for Applied Life Sciences, Amherst Life Sciences Laboratories, University of Massachusetts , 240 Thatcher Road, Amherst, MA 01003

Abstract

Abstract Our objective was to develop a technique for performing irreversible electroporation (IRE) of esophageal tumors while mitigating thermal damage to the healthy lumen wall. We investigated noncontact IRE using a wet electrode approach for tumor ablation in a human esophagus with finite element models for electric field distribution, joule heating, thermal flux, and metabolic heat generation. Simulation results indicated the feasibility of tumor ablation in the esophagus using an catheter mounted electrode immersed in diluted saline. The ablation size was clinically relevant, with substantially lesser thermal damage to the healthy esophageal wall when compared to IRE performed by placing a monopolar electrode directly into the tumor. Additional simulations were used to estimate ablation size and penetration during noncontact wet-electrode IRE (wIRE) in the healthy swine esophagus. A novel catheter electrode was manufactured and wIRE evaluated in seven pigs. wIRE was performed by securing the device in the esophagus and using diluted saline to isolate the electrode from the esophageal wall while providing electric contact. Computed tomography and fluoroscopy were performed post-treatment to document acute lumen patency. Animals were sacrificed within four hours following treatment for histologic analysis of the treated esophagus. The procedure was safely completed in all animals; post-treatment imaging revealed intact esophageal lumen. The ablations were visually distinct on gross pathology, demonstrating full thickness, circumferential regions of cell death (3.52 ± 0.89 mm depth). Acute histologic changes were not evident in nerves or extracellular matrix architecture within the treatment site. Catheter directed noncontact IRE is feasible for performing penetrative ablations in the esophagus while avoiding thermal damage.

Funder

Congressionally Directed Medical Research Programs

National Cancer Institute

National Institute of Diabetes and Digestive and Kidney Diseases

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering