Optimal Design of a Medical Staple Using Shape Memory Alloy

Author:

Lin Y.1,Huang M.1,Zhang M.1

Affiliation:

1. Department of Mechanical & Aerospace Engineering, University of Missouri, Columbia, MO 65211

Abstract

This paper presents the mechanical design of a medical staple using shape memory alloy (SMA) material. The design problem was formulated as an optimization problem minimizing an energy functional. The constraints were the length of wire, the properties of the SMA material, and the required clamping force. The Euler-Lagrange equation of this isoperimetric calculus of variation problem was derived, which was a fourth order two-point boundary value problem. An analytical solution was constructed and verified. A numerical optimization technique was applied to generate other possible solutions for the boundary value problem. The objective was to minimize the error of the solution for specific boundary conditions. The sequential quadratic programming algorithm was applied with Runge-Kutta integration to solve the boundary value problem. The design procedure described in this paper could be applied to the design of other mechanical components of wire form using SMA material.

Publisher

ASME International

Subject

Computer Graphics and Computer-Aided Design,Computer Science Applications,Mechanical Engineering,Mechanics of Materials

Reference29 articles.

1. Assad M. , LombardiS., BernecheS., DesrosiersE. A., YahiaL. H., and RivardC. H., 1994, “Assays of Cytotoxicity of the Nickel-Titanium Shape Memory Alloy,” Annales de Chirurgie, Vol. 48, No. 8, pp. 7311–736. (In French)

2. Bolza, O., 1961, Lectures on the calculus of variations, Dover publications.

3. Castleman, L. S., 1973, The 5th Annual Biomaterials Symposium, Clemson, S.C.

4. Cutright, D. E., Bhaskar, S. N., Perez, B., Johnson, R. M., and Cowan, G. S., 1973, Oral Surg., Vol. 35, No. 4.

5. Dai K. R. , 1983, “Application of NiTi Shape Memory Alloy in Double-Cup Prosthesis of Hip,” Chinese Journal of Surgery, Vol. 21, No. 9, pp. 540–541. (In Chinese)

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Measuring mechanical properties of fine-wire cross-sections used in medical devices;Journal of Biomedical Materials Research;2004

2. ELASTIC BUCKLING OF EXTERNALLY-PRESSURIZED IMPERFECT TOROIDAL SHELLS;International Journal of Structural Stability and Dynamics;2001-03