Hot-Film Wall Shear Probe Measurements Inside a Ventricular Assist Device

Author:

Baldwin J. T.1,Tarbell J. M.2,Deutsch S.3,Geselowitz D. B.1,Rosenberg G.4

Affiliation:

1. Bioengineering Program, The Pennsylvania State University, University Park, Pa. 16802

2. Department of Chemical Engineering, The Pennsylvania State University, University Park, Pa. 16802

3. Applied Research Laboratory, The Pennsylvania State University, University Park, Pa. 16802

4. Division of Artificial Organs, Department of Surgery, The Pennsylvania Park University, Hershey, Pa. 17033

Abstract

Wall shear rates at eleven sites within the Penn State Electric Ventricular Assist Device (EVAD) were determined with the pump operating under conditions of 30 and 50 percent systolic duration and a mean flow rate of 5.8 L/min using a flush-mounted hot-film probe. Probe calibrations were performed with the hot-film in two orientations relative to the flow direction: a standard orientation and an orientation in which the hot-film was rotated by 90 deg from the standard orientation. The magnitude and direction of the wall shear stress at each site within the EVAD were estimated from ensemble averaged voltage data recorded for similar standard and rotated film orientations. The results indicate that, during diastole the wall shear stress direction around the pump’s periphery for both operating conditions is predominantly perpendicular to the inflow-outflow plane (in the direction of the pusher plate motion) and reaches a peak value of approximately 350 dynes/cm2. The highest wall shear stresses were found near the prosthetic aortic valve (inside the EVAD) under the 30 percent systolic duration condition and are estimated to be as high as 2700 dynes/cm2. Peak shear stress values of 1400 dynes/cm2 were observed in the vicinity of the prosthetic mitral valve under both operating conditions. The results suggested that the valve regions are substantially more hemolytic than other wall regions of the EVAD; the magnitudes of the wall shear stresses are sensitive to operating conditions; and that wall shear in the direction of pusher plate motion can be significant.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Cited by 21 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Analysis of cavitation and shear in bellows pump: transient CFD modelling and high-speed visualization;Engineering Applications of Computational Fluid Mechanics;2023-08-19

2. John M. Tarbell: Artificial Heart and Mechanical Heart Valve Research Contributions;Cardiovascular Engineering and Technology;2021-01-06

3. Flow Measurement;Handbook of Measurement in Science and Engineering;2013-01-04

4. Evaluating the Flow Characteristics of Artificial Pumping Devices Using Nuclear Scintigraphsy;Artificial Organs;2008-11-12

5. UNSTEADY FLOW IN A 3D CENTRALLY-ACTUATED TYPE ARTIFICIAL HEART;International Journal of Modern Physics C;2007-05

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3