Total Knee Replacement Polyethylene Stresses During Loading in a Knee Simulator
Author:
Giddings Virginia L.1, Kurtz Steven M.1, Edidin Avram A.2
Affiliation:
1. Exponent, Inc., 149 Commonwealth Avenue, Menlo Park, CA 94025 2. Howmedica Osteonics Corp., 59 Route 17, Allendale, NJ 07401
Abstract
We investigated the stresses and kinematics of a total knee replacement during the duty cycle of a knee simulator. Finite element models were constructed of the tibial and the femoral component of a commercially available cruciate retaining total knee replacement. Time dependent flexion/extension, axial loading, and anterior/posterior loading were applied to the components of the arthroplasty to match those generated by the knee simulator. We evaluated the effect of varying the stiffness of a spring-loaded bumper system for anterior-posterior constraint on the joint kinematics as well as on the stresses within the polyethylene tibial component. Both the joint kinematics and the stresses and strains subjected to the polyethylene tibial component, were found to be comparatively insensitive to the stiffness of the spring bumper system for this design. When the stiffness of the bumper system was increased by two orders of magnitude, the maximum contact stresses, von Mises stresses, and von Mises strains in the polyethylene tibial component varied by only 15 to 59 percent. In general, increasing the stiffness of the bumper system decreased the displacements of the base plate, but the relationships were nonlinear, possibly due to the added constraints imposed by the tibiofemoral contact interaction. The long-term goal of this research is to develop a validated structural model to predict the stresses, kinematics, and ultimately, the wear, of total joint replacement components in a contemporary knee joint simulator.
Publisher
ASME International
Subject
Surfaces, Coatings and Films,Surfaces and Interfaces,Mechanical Engineering,Mechanics of Materials
Reference19 articles.
1. Hood, R. W., Wright, T. M., and Burstein, A. H., 1983, “Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses,” J. Biomed. Mater. Res., 17, No. 5, pp. 829–842. 2. Schmalzried, T. P., and Callaghan, J. J., 1999, “Wear in total hip and knee replacements,” J. Bone Jt. Surg., Am., 81, No. 1, pp. 115–36. 3. Peters, P. C., Jr., Engh, G. A., Dwyer, K. A., and Vinh, T. N., 1992, “Osteolysis after total knee arthroplasty without cement,” J. Bone Jt. Surg., Am., 74, No. 6, pp. 864–876. 4. Wright, T. M., and Bartel, D. L., 1986, “The problem of surface damage in polyethylene total knee components,” Clin. Orthop., 205, pp. 67–74. 5. Bartel, D. L., Bicknell, V. L., and Wright, T. M., 1986, “The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement,” J. Bone Jt. Surg., Am., 68, pp. 1041–1051.
Cited by
25 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|