Simplified Multistage Computational Approach to Assess the Fatigue Behavior of a Niti Transcatheter Aortic Valve During In Vitro Tests: A Proof-of-Concept Study

Author:

Petrini Lorenza1,Dordoni Elena2,Allegretti Dario2,Pott Desiree3,Kütting Maximilian3,Migliavacca Francesco2,Pennati Giancarlo4

Affiliation:

1. Department of Civil and Environmental Engineering, Politecnico di Milano, Milano 20133, Italy

2. Laboratory of Biological Structure Mechanics, Department of Chemistry, Materials and Chemical Engineering “Giulio Natta,” Politecnico di Milano, Milano 20133, Italy

3. Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Helmholtz Institute, RWTH Aachen University, Aachen 52062, Germany

4. Laboratory of Biological Structure Mechanics, Department of Chemistry, Materials and Chemical Engineering “Giulio Natta,” Politecnico di Milano, Piazza Leonardo da Vinci 32, Milano 20133, Italy e-mail:

Abstract

Nowadays, transcatheter aortic valve (TAV) replacement is an alternative to surgical therapy in selected high risk patients for the treatment of aortic stenosis. However, left ventricular contraction determines a severe cyclic loading for the implanted stent-frame, undermining its long-term durability. Technical standards indicate in vitro tests as a suitable approach for the assessment of TAV fatigue behavior: generally, they do not specify test methods but require to test TAV in the worst loading conditions. The most critical conditions could be different according to the specific valve design, hence the compartment where deploying the valve has to be properly identified. A fast and reliable computational methodology could significantly help to face this issue. In this paper, a numerical approach to analyze Nickel-Titanium TAV stent-frame behavior during in vitro durability tests is proposed. A simplified multistage strategy was adopted where, in each stage, only two of the three involved components are considered. As a proof-of-concept, the method was applied to a TAV prototype. Despite its simplifications, the developed computational framework gave useful insights into the stent-frame failures behavior during a fatigue test. Numerical results agree with experimental findings. In particular, the most dangerous condition was identified among a number of experimental tests, where different compartments and pressure gradients were investigated. The specific failure location was also correctly recognized. In conclusion, the presented methodology provides a tool to support the choice of proper testing conditions for the in vitro assessment of TAV fatigue behavior.

Funder

State of North Rhine-Westphalia

Publisher

ASME International

Subject

Biomedical Engineering,Medicine (miscellaneous)