Local Head Roughening as a Factor Contributing to Variability of Total Hip Wear: A Finite Element Analysis
Author:
Brown Thomas D.1, Stewart Kristofer J.1, Nieman John C.2, Pedersen Douglas R.1, Callaghan John J.1
Affiliation:
1. Department of Orthopaedic Surgery, University of Iowa, Iowa City, IA 52242 2. Department of Biomedical Engineering, University of Iowa, Iowa City, IA 52242
Abstract
Large inter-patient variability in wear rate and wear direction have been a ubiquitous attribute of total hip arthroplasty (THA) cohorts. Since patients at the high end of the wear spectrum are of particular concern for osteolysis and loosening, it is important to understand why some individuals experience wear at a rate far in excess of their cohort average. An established computational model of polyethylene wear was used to test the hypothesis that, other factors being equal, clinically typical variability in regions of localized femoral head roughening could account for much of the variability observed clinically in both wear magnitude and wear direction. The model implemented the Archard abrasive/adhesive wear relationship, which incorporates contact stress, sliding distance, and (implicitly) bearing surface tribology. Systematic trials were conducted to explore the influences of head roughening severity, roughened area size, and roughened area location. The results showed that, given the postulated wear factor elevations, head roughening variability (conservatively) typical of retrieval specimens led to approximately a 30° variation in wear direction, and approximately a 7-fold variation in volumetric wear rate. Since these data show that randomness in head scratching can account for otherwise-difficult-to-explain variations in wear direction and wear rate, third-body debris may be a key factor causing excessive wear in the most problematic subset of the THA population.
Publisher
ASME International
Subject
Physiology (medical),Biomedical Engineering
Reference45 articles.
1. Charnley, J., and Halley, D. K., 1975, “Rate of Wear in Total Hip Replacement,” Clin. Orthop., 112, pp. 170–179. 2. Clarke, J. C., Black, K., Rennie, C., and Amstutz, H. C., 1976, “Can Wear in Total Hip Arthroplastics Be Assessed from Radiographs?,” Clin. Orthop., 121, pp. 126–142. 3. Pedersen, D. R., Brown, T. D., Hillis, S. L., and Callaghan, J. J., 1998, “Prediction of Long-Term Polyethylene Wear in Total Hip Arthroplasty, Based on Early Wear Measurements Made Using Digital Image Analysis,” J. Orthop. Res., 16(5), pp. 557–563. 4. Schmalzried, T. P., Dorey, F. J., and McKellop, H., 1998, “The Multifactorial Nature of Polyethylene Wear in vivo,” J. Bone Jt. Surg., Am. Vol., 80(8), pp. 1234–1242. 5. Devane, P. A., Robinson, E. J., Bourne, R. B., Rorabeck, C. H., Nayak, N. N., and Horne, J. G., 1997, “Measurement of Polyethylene Wear in Acetabular Components Inserted with and without Cement. A Randomized Trial,” J. Bone Jt. Surg., Am. Vol., 79(5), pp. 682–689.
Cited by
43 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|