A Precision System for Computed Tomography-Guided Needle Placement in the Thorax and Abdomen—Technical Design and Performance Analysis

Author:

Arnolli Maarten M.12,Buijze Martijn3,Franken Michel3,Broeders Ivo A. M. J.4,Brouwer Dannis M.5

Affiliation:

1. DEMCON Advanced Mechatronics, Enschede 7521 PH, The Netherlands;

2. Precision Engineering Science-Based Engineering, University of Twente, Enschede 7522 NB, The Netherlands e-mail:

3. DEMCON Advanced Mechatronics, Enschede 7521 PH, The Netherlands

4. Robotics and Mechatronics, MIRA Institute for Biomedical Technology and Technical Medicine, University of Twente, Enschede 7522 NB, The Netherlands

5. Precision Engineering Science-Based Engineering, University of Twente, Enschede 7522 NB, The Netherlands

Abstract

A system was developed for computed tomography (CT)-guided needle placement in the thorax and abdomen, providing precise aiming of a needle guide (NG) to reach a user-specified target in a single manual insertion. The objective of this work is to present its technical design and analyze its performance in terms of placement error in air. The individual contributions to the placement error of a fiducial marker based system-to-CT registration system, a two degrees-of-freedom (2DOFs) drive system to aim the NG, and a structural link between NG and CT table were experimentally determined, in addition to the placement error of the overall system. An error contribution of 0.81 ± 0.34 mm was determined for the registration system, <1.2 mm and <3.3 mm for the drive system, and 0.35 mm and 0.43 mm for two load cases of the structural link. The overall unloaded system achieved 1.0 ± 0.25 mm and 2.6 ± 0.7 mm at 100 mm and 250 mm depth, respectively. The overall placement errors in air do not exceed the ≤5 mm error specified as a clinical user requirement for needle placement in tissue.

Funder

ITEA3

Publisher

ASME International

Subject

Biomedical Engineering,Medicine (miscellaneous)

Reference11 articles.

1. Biopsy of the Liver;J.,2011

2. Percutaneous Ablation of Hepatic Tumors;Semin. Interventional Radiol.,2010

3. Percutaneous Lung Biopsy: Technique, Efficacy, and Complications;Semin. Interventional Radiol.,2013

4. Current Status of Thermal Ablation Treatments for Lung Malignancies;Semin. Interventional Radiol.,2010

5. An Overview of Systems for CT- and MRI-Guided Percutaneous Needle Placement in the Thorax and Abdomen;Int. J. Med. Rob. Comput. Assisted Surg.,2015

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Robotic systems in interventional oncology: a narrative review of the current status;International Journal of Clinical Oncology;2023-04-28

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3