In Vitro Evaluation of the Acetabular Cup Primary Stability by Impact Analysis

Author:

Michel Adrien1,Bosc Romain2,Vayron Romain1,Haiat Guillaume3

Affiliation:

1. Laboratoire Modélisation et de Simulation Multi-Echelle, CNRS, UMR CNRS 8208, 61 Avenue du Général de Gaulle, Créteil 94010, France

2. Service de Chirurgie Plastique, Reconstructrice et Esthétique, Hôpital Henri Mondor AP-HP, CHU Paris 12, Université Paris-Est, 51 Avenue du Maréchal de Lattre de Tassigny, Créteil 94000, France

3. Laboratoire Modélisation et de Simulation Multi-Echelle, CNRS, UMR CNRS 8208, 61 Avenue du Général de Gaulle, Créteil 94010, France e-mail:

Abstract

The implant primary stability of the acetabular cup (AC) is an important parameter for the surgical success of press-fit procedures used for the insertion of cementless hip prostheses. In previous studies by our group (Mathieu, V., Michel, A., Lachaniette, C. H. F., Poignard, A., Hernigou, P., Allain, J., and Haiat, G., 2013, “Variation of the Impact Duration During the in vitro Insertion of Acetabular Cup Implants,” Med. Eng. Phys., 35(11), pp. 1558–1563) and (Michel, A., Bosc, R., Mathieu, V., Hernigou, P., and Haiat, G., 2014, “Monitoring the Press-Fit Insertion of an Acetabular Cup by Impact Measurements: Influence of Bone Abrasion,” Proc. Inst. Mech. Eng., Part H, 228(10), pp. 1027–1034), the impact momentum and duration were shown to carry information on the press-fit insertion of the AC within bone tissue. The aim of the present study is to relate the impact momentum recorded during the AC insertion to the AC biomechanical primary stability. The experimental protocol consisted in testing 13 bovine bone samples that underwent successively series of 15 reproducible mass falls impacts (5 kg, 5 cm) followed by tangential stability testing. Each bone sample was tested with different hole sizes in order to obtain different stability configurations. The impact momentum and the tangential primary stability reach a maximum value for an interference fit equal to around 1 mm. Moreover, a correlation between the impact momentum and the stability was obtained with all samples and all configuration (R2 = 0.65). The implant primary stability can be assessed through the measurement of the impact force signal analysis. This study opens new paths for the development of a medical device which could be used as a decision support system to assist the surgeon during the insertion of the AC implant.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference42 articles.

1. Cementless Press-Fit Cup. Principles, Experimental Data, and Three-Year Follow-Up Study;Clin. Orthop. Relat. Res.,1989

2. The Effect of Under-Reaming on the Cup/Bone Interface of a Press Fit Hip Replacement;ASME J. Biomech. Eng.,2010

3. Factors Influencing Stability at the Interface Between a Porous Surface and Cancellous Bone: A Finite Element Analysis of a Canine In Vivo Micromotion Experiment;J. Biomed. Mater. Res.,1997

4. Biomechanical Determinants of the Stability of Dental Implants: Influence of the Bone-Implant Interface Properties;J. Biomech.,2014

5. Observations on the Effect of Movement on Bone Ingrowth Into Porous-Surfaced Implants;Clin. Orthop. Relat. Res.,1986

Cited by 23 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3