Modification of a Simplified Hip Joint Simulator Into an ISO 14242-1 Compliant Design and a Comparison of Wear Test Results

Author:

Saikko Vesa1,Morad Omar1,Viitala Raine1

Affiliation:

1. Department of Mechanical Engineering, Aalto University School of Engineering, P. O. Box 14400, Aalto FI-00076, Finland

Abstract

Abstract The ISO 14242-1 standard for hip joint simulator wear testing specifies a set of test conditions for the simulation of normal level walking in optimal conditions. Since some of the established simulators, such as the 12-station HUT-4, are not ISO 14242-1 compliant, the present study was carried out to answer the following question. Does wear produced in ISO 14242-1 conditions differ from that obtained earlier with the simplified HUT-4 hip joint simulator for similar specimens in normal level walking, optimal conditions? The established HUT-4 hip joint simulator was made ISO compliant by an implementation of a number of modifications. One of the modifications was the design and construction of a novel servo-electric load actuator with a proven dynamic response. The other modifications were related to the Euler sequence of motions, acetabular abduction angle, enclosure of the lubricant chamber, and temperature control. A 5 million-cycle wear test with thin, large-diameter VEXLPE liners resulted in a wear rate close to that obtained earlier with the HUT-4. The burnished bearing surface in both tests was in agreement with clinical retrieval studies. It appeared that a more simple, inexpensive hip joint simulator can reproduce clinical wear mechanisms. However, the simulator must meet certain basic requirements. The most important of these is likely to be the correct type of multidirectional relative motion, for which biaxial motion is sufficient. It is emphasized that the present study was not intended to show a similarity in wear produced by the ISO 14242-1 and HUT-4 wear test systems.

Publisher

ASME International

Subject

Surfaces, Coatings and Films,Surfaces and Interfaces,Mechanical Engineering,Mechanics of Materials

Reference32 articles.

1. The Origin of Submicron Polyethylene Wear Debris in Total Hip Arthroplasty;McKellop;Clin. Orthop. Relat. Res.,1995

2. Force Track Analysis of Contemporary Hip Simulators;Calonius;J. Biomech.,2003

3. A 12-Station Anatomic Hip Joint Simulator;Saikko;Proc. Inst. Mech. Eng. H.,2005

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3