Design and Analysis of Orthogonally Compliant Features for Local Contact Pressure Relief in Transtibial Prostheses

Author:

Faustini Mario C.1,Crawford Richard H.1,Neptune Richard R.1,Rogers William E.2,Bosker Gordon2

Affiliation:

1. Department of Mechanical Engineering, The University of Texas at Austin, Austin, Texas 78712

2. Department of Rehabilitation Medicine, The University of Texas Health Science Center at San Antonio, San Antonio, Texas 78229

Abstract

A very attractive advantage of manufacturing prosthetic sockets using solid freeform fabrication is the freedom to introduce design solutions that would be difficult to implement using traditional manufacturing techniques. Such is the case with compliant features embedded in amputee prosthetic sockets to relieve contact pressure at the residual limb-socket interface. The purpose of this study was to present a framework for designing compliant features to be incorporated into transtibial sockets and manufacturing prototypes using selective laser sintering (SLS) and Duraform™ material. The design process included identifying optimal compliant features using topology optimization algorithms and integrating these features within the geometry of the socket model. Using this process, a compliant feature consisting of spiral beams and a supporting external structure was identified. To assess its effectiveness in reducing residual limb-socket interface pressure, a case study was conducted using SLS manufactured prototypes to quantify the difference in interface pressure while a patient walked at his self-selected pace with one noncompliant and two different compliant sockets. The pressure measurements were performed using thin pressure transducers located at the distal tibia and fibula head. The measurements revealed that the socket with the greatest compliance reduced the average and peak pressure by 22% and 45% at the anterior side distal tibia, respectively, and 19% and 23% at the lateral side of the fibula head, respectively. These results indicate that the integration of compliant features within the socket structure is an effective way to reduce potentially harmful contact pressure and increase patient comfort.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference15 articles.

1. Fabrication of Prosthetic Sockets by Selective Laser Sintering;Rogers

2. A Computerized System to Manufacture Prostheses for Amputees in Developing Countries;Walsh;Prosthet. Orthot Int.

3. Faustini, M. , Modeling and Fabrication of Prosthetic Sockets using Selective Laser Sintering, Doctoral dissertation, The University of Texas at Austin, 2004.

4. State-of-the-Art Research in Lower-Limb Prosthetic Biomechanics-Socket Interface;Mak;J. Rehabil. Res. Dev.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3