MRI-Based Modeling for Radiocarpal Joint Mechanics: Validation Criteria and Results for Four Specimen-Specific Models

Author:

Fischer Kenneth J.,Johnson Joshua E.1,Waller Alexander J.2,McIff Terence E.,Bruce Toby E.3,Bilgen Mehmet4

Affiliation:

1. Department of Mechanical Engineering, University of Kansas, 1530 W 15th Street, Room 3138,Lawrence, KS 66045-7609

2. Midwest Research Institute, 425 Volker Boulevard,Kansas City, MO 64110e-mail:

3. Department of Orthopedic Surgery, University of Kansas Medical Center, 3901 Rainbow Blvd., Mail Stop 3017, Kansas City, KS 66160-7387

4. Biomedical Imaging Department,Faculty of Medicine,University of Malaysia, Kuala Lumpur, 50603 Malaysia e-mail:

Abstract

The objective of this study was to validate the MRI-based joint contact modeling methodology in the radiocarpal joints by comparison of model results with invasive specimen-specific radiocarpal contact measurements from four cadaver experiments. We used a single validation criterion for multiple outcome measures to characterize the utility and overall validity of the modeling approach. For each experiment, a Pressurex film and a Tekscan sensor were sequentially placed into the radiocarpal joints during simulated grasp. Computer models were constructed based on MRI visualization of the cadaver specimens without load. Images were also acquired during the loaded configuration used with the direct experimental measurements. Geometric surface models of the radius, scaphoid and lunate (including cartilage) were constructed from the images acquired without the load. The carpal bone motions from the unloaded state to the loaded state were determined using a series of 3D image registrations. Cartilage thickness was assumed uniform at 1.0 mm with an effective compressive modulus of 4 MPa. Validation was based on experimental versus model contact area, contact force, average contact pressure and peak contact pressure for the radioscaphoid and radiolunate articulations. Contact area was also measured directly from images acquired under load and compared to the experimental and model data. Qualitatively, there was good correspondence between the MRI-based model data and experimental data, with consistent relative size, shape and location of radioscaphoid and radiolunate contact regions. Quantitative data from the model generally compared well with the experimental data for all specimens. Contact area from the MRI-based model was very similar to the contact area measured directly from the images. For all outcome measures except average and peak pressures, at least two specimen models met the validation criteria with respect to experimental measurements for both articulations. Only the model for one specimen met the validation criteria for average and peak pressure of both articulations; however the experimental measures for peak pressure also exhibited high variability. MRI-based modeling can reliably be used for evaluating the contact area and contact force with similar confidence as in currently available experimental techniques. Average contact pressure, and peak contact pressure were more variable from all measurement techniques, and these measures from MRI-based modeling should be used with some caution.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference37 articles.

1. A Framework for the In Vivo Pathomechanics of Osteoarthritis at the Knee;Andriacchi;Ann. Biomed. Eng.

2. Experimental Arthritis in Rabbit Knees: A Study of Relief of Pressure on One Tibial Plateau in Immature and Mature Rabbits;Engh;Clin. Orthop. Relat. Res.

3. Cartilage Changes after Experimental Relief of Contact in the Knee Joint of the Mature Rat;Hall;Clin. Orthop. Relat. Res.

4. A Biomechanical Analysis of Joint Contact Forces in the Posterior Cruciate Deficient Knee;MacDonald;Knee Surg. Sports Traumatol. Arthrosc.

5. Biomechanics of the Wrist;Patterson;J. Hand Ther.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3