The Development of a Novel Peristaltic Test Stand “Swallow-Sim” for the Mechanical Evaluation of Esophageal Stents

Author:

Berger Daniel Michael12ORCID,Sehrt Jan T.3,Brinkmann Thomas4,Henrichs Niclas Norman4ORCID,Bilec Oguzhan4,Fuchs Karl-Hermann5,Meining Alexander5

Affiliation:

1. Impetus Plastics Engineering GmbH, Bahnhofstraße 9 , Bad Aibling 83043, Germany ; , Universitätsstraße 150, Bochum 44780, Germany

2. Product Development, Ruhr-Universität Bochum , Bad Aibling 83043, Germany ; , Universitätsstraße 150, Bochum 44780, Germany

3. Hybrid Additive Manufacturing, Ruhr-Universität Bochum , Universitätsstraße 150, Bochum 44780, Germany

4. Impetus Plastics Engineering GmbH , Bahnhofstraße 9, Bad Aibling 83043, Germany

5. Gastroenterology, Universitätsklinikum Würzburg , Josef-Schneider-Straße 2, Würzburg 97070, Germany

Abstract

Abstract The development of medical products begins with the “in silico” phase, where the development and simulation of new stent types are carried out. This is followed by the “in vitro” phase. Here, tests are done in a test stand to obtain initial conclusions about the interaction of the environment. The approval process is completed in the “in vivo” phase, where testing in living beings happen. Here, preclinical studies are carried out in animals first, followed by clinical studies on patients. A big part of the development fails in this final phase, as this is where the interactions of all influences from the stent environment are investigated. Since this not only causes high costs for the developers but also unnecessarily destroys living resources in animal studies, this publication describes the development of a test stand called “Swallow-Sim” that superimposes the mechanical influences of the esophagus, the chemical stress caused by hydrochloric acid and increased body temperature. Furthermore, tests of the acting pressures are carried out using esophageal manometry, a temperature test of the test stand and a test run of the gastric juice. At the end of this publication, the results are evaluated with a six-week test of a Nickel Titanium Naval Ordnance Laboratory stent, which loses much of its mechanical properties and is partially destroyed by the load. The results show a clear correlation with the findings from reality. The test stand should be further optimized and examined in more detail in further tests and subjected to a reality check.

Publisher

ASME International

Reference52 articles.

1. Krebs in Deutschland Für 2019/2020—C13 Speiseröhrenkrebs;Robert Koch Institute,2023

2. Recent Advancements in Esophageal Cancer Treatment in Japan;Ann. Gastroenterol. Surg.,2018

3. Development of a Prediction Model of Adverse Events After Stent Placement for Esophageal Cancer;Gastrointest. Endosc.,2016

4. The Role of Endoscopy in the Assessment and Treatment of Esophageal Cancer;Gastrointest. Endosc.,2013

5. Esophageal Dilation;Gastrointest. Endosc.,2006