Author:
Herrling P.,Zürn A.,Anders P.,Kotzerke J.,Wunderlich G.
Abstract
SummaryThin layer chromatography is well established for quality control of radiopharmaceuticals. A convenient and widely used stationary phase are ITLC SG strips. However, the Pall Corporation stopped manufacturing of the silica gel impregnated glass fibre strips (ITLC SG). Material, Methode: As a replacement we tested silicic acid impregnated glass fibre strips from Varian (ITLC SA) and sufficient mobile phases. Results: The chromatography with these strips takes two to three times longer than with ITLC SG, but it is in an acceptable range. Only three mobile phases are necessary to test most of the common in-house made radiopharmaceuticals. Conclusion: The proposed method is suitable for routinely measuring the radiochemical purity of radiophamaceuticals.
Subject
Radiology Nuclear Medicine and imaging,General Medicine
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献