Abstract
While the sampling of pharmaceutical products typically follows well-defined protocols, the parameterization of spectroscopic methods and their associated sampling frequency is not standard. Whereas, for blending, the sampling frequency is limited by the nature of the process, in other processes, such as tablet film coating, practitioners must determine the best approach to collecting spectral data. The present article studied how sampling practices affected the interpretation of the results provided by a near-infrared spectroscopy method for the monitoring of tablet moisture and coating weight gain during a pan-coating experiment. Several coating runs were monitored with different sampling frequencies (with or without co-adds (also known as sub-samples)) and with spectral averaging corresponding to processing cycles (1 to 15 pan rotations). Beyond integrating the sensor into the equipment, the present work demonstrated that it is necessary to have a good sense of the underlying phenomena that have the potential to affect the quality of the signal. The effects of co-adds and averaging was significant with respect to the quality of the spectral data. However, the type of output obtained from a sampling method dictated the type of information that one can gain on the dynamics of a process. Thus, different sampling frequencies may be needed at different stages of process development.
Subject
Spectroscopy,Instrumentation