Affiliation:
1. Institute of Analytical and Bioanalytical Chemistry, University of Ulm, Ulm, Germany
Abstract
The analytical performance of four commercially available infrared attenuated total reflection (IR-ATR) accessories with various ATR waveguide materials has been analyzed and evaluated using acetate, CO2, and CO32− solutions. Calibration functions have been established to determine and compare analytically relevant parameters such as sensitivity, signal-to-noise ratio (SNR), and efficiency. The obtained parameters were further analyzed to support conclusions on the differences in performance of the individual IR-ATR accessories.
Subject
Spectroscopy,Instrumentation