Affiliation:
1. Department of Chemical Pathology, Bristol Royal Infirmary, Marlborough Street, Bristol BS2 8HW, UK
2. Accident and Emergency Department, Bristol Royal Infirmary, Marlborough Street, Bristol BS2 8HW, UK
Abstract
The i-STAT© point-of-care testing analyser, which utilizes single-use multisensor cartridges, was evaluated. Two cartridges were assessed: the 6+ and the G3 +, which provide results for urea, glucose, sodium, potassium, chloride, haematocrit, pH, PCO2, PO2 and various calculated parameters. The results for all analytes agreed well with the analysers in routine use in the laboratory. The reproducibility was comparable even when analysis was carried out by a nurse with only 5 min training. The system was found to be reliable, easy to use and required no maintenance (only a 2-min daily check of the electronics). These features, together with portability and the storage capacity for results, make the i-STAT suitable for point-of-care use, particularly in critical care units.
Subject
Clinical Biochemistry,General Medicine
Cited by
50 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献