Affiliation:
1. Endocrine Unit, Clinical Biochemistry Department, Oldchurch Hospital, Romford, Essex RM7 0BE, UK
Abstract
A microtiter plate based Dual Analyte enzymeimmunoassay method for the simultaneous measurement of α-fetoprotein (AFP) and Free-β human chorionic gonadotrophin (hCG) was evaluated. This rapid assay, which has application in both Neural Tube Defect screening and Down's screening, shows good precision with between assay coefficients of variation between 5 and 7·5% for AFP and 3·7 to 5·8% for Free-β(hCG). Correlation with single analyte procedures is good, with correlation coefficients being greater than 0·91 in both cases. Clinical discrimination in detecting both types of abnormalities is not compromised by this new simultaneous Dual Analyte assay. We conclude that the Dual Analyte approach, which combines analytes achieving the highest known detection efficiency, will bring about improvements in the efficiency of screening, reduce costs and improve report turnaround, all leading to better quality of patient care.
Subject
Clinical Biochemistry,General Medicine
Cited by
27 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献