Affiliation:
1. Clinical Laboratory, Princess Irene Hospital, Almelo, The Netherlands
Abstract
Urinary cobalt was determined by flameless atomic absorption spectroscopy. Three methods were compared: (i) direct analysis with deuterium background correction after 11-fold dilution with distilled water (method D), (ii) analysis with deuterium background correction after extraction of cobalt from the urinary matrix in organic solution (method E), and (iii) direct analysis with Zeeman background correction (method Z). The detection limit of the direct analysis of urinary cobalt with deuterium background correction was 6 μg/L, this appeared to be insufficient for the determination of reference values and moderate enlarged cobalt values. Comparison of the extraction method and the Zeeman method revealed that these methods have similar reference values, detection limits, precision and recovery.
Subject
Clinical Biochemistry,General Medicine
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献