Affiliation:
1. Technicon Methods and Standards Laboratory, 54-58 Bartholomew Close, London EC1A 7HL
Abstract
An automated assay for serum amylase is described. Using the AutoAnalyzer II the method employs, as substrate, a solution of dyed amylopectin which is commercially available (DyAmyl-L). After incubation of serum and substrate at 55°C, 1·0 mol/l KOH is introduced and the coloured products of enzymic hydrolysis are separated by dialysis against phosphate buffer. Absorbance is measured at 540 nm and the enzymic activity read from a standard curve obtained using calibration sera Versatol E and EN. The system provides accurate and reproducible measurement of serum amylase. The results obtained from assaying 100 ‘normal’ sera are compared with those using another method of α-amylase estimation.
Subject
Clinical Biochemistry,General Medicine
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献