Affiliation:
1. Department of Chemical Pathology, Liverpool Area Health Authority (Teaching), Ashton Street, Liverpool L3 5RT
Abstract
An automated method is described, using standard continuous flow techniques, for the determination of urine fluoride ion concentration using a fluoride ion selective electrode. It is shown that the kinetics of the electrode response to changes in fluoride ion can be used for the accurate measurement of fluoride ion concentration in urine, and that equilibration of the electrode response is not a prerequisite for the measurement of fluoride ion. Recovery experiments are in the range 83 to 90%; in-batch precision is between 0·9 and 1·6% and carryover 2·5% or less.