Affiliation:
1. Department of Laboratory Science, the Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
Abstract
Haemoglobinopathies may interfere with the haemoglobin A1c (HbA1c) measurement, leading to incorrect diagnosis and inappropriate treatment. It is essential that HbA1c assays are capable of identifying haemoglobinopathies. We report two cases of haemoglobin New York (HbNY) discovered through HbA1c analysis using capillary electrophoresis (Capillarys 2 Flex Piercing [C2FP], Sebia). We used these samples to evaluate the ability of three other HbA1c assays to identify this variant: ion-exchange high-performance liquid chromatography (Variant II Turbo [VII-T], Bio-Rad); boronate affinity high-performance liquid chromatography (Ultra2, Trinity Biotech) and immunoassay (Cobas c501 Tina-quant Generation 3, Roche Diagnostics). Each method was used for HbA1c assay of in samples from two cases of heterozygous haemoglobinopathy: β0-thalassemia/HbNY (Case 1) and HbA/NY (Case 2). Only the C2FP system detected HbNY (an additional peak appeared between HbA1c and HbA0). Clinical laboratories should be aware of the limitations of their HbA1c assay methods especially in geographic areas, where haemoglobinopathy prevalence is high.
Subject
Clinical Biochemistry,General Medicine
Cited by
1 articles.