The interference by HbF on HbA1c (BM Test HbA1c) measurement in enzymatic method

Author:

Nitta Takenori1,Yamashiro Yasuhiro1,Hattori Yukio1,Ezumi Tomoko2,Nishioka Mituaki2,Nakamura Junji2

Affiliation:

1. Department of Health Science, Yamaguchi University Graduate School of Medicine, Ube, Japan

2. Department of Clinical Laboratory, Yamaguchi University Hospital, Ube, Japan

Abstract

Background High performance liquid chromatography and immunological and enzymatic methods are known as the methods for HbA1c measurement. However, the differences in the principles of the methods may cause slight discrepancies, which become problems especially in the regions where hemoglobinopathies that have high HbF concentrations are commonly seen. In this report, we discuss the effect of HbF on the HbA1c measurement by the enzymatic method (BM Test HbA1c) which can be applied to automatic analyzer of high sample throughput. Methods Thirty-five samples (HbF concentration: between 1.4 and 90.1%) and 68 samples (HbF concentration: <1.0%) were analyzed by high performance liquid chromatography (Tosoh G8) and BM Test HbA1c. The results by BM Test HbA1c were compared with those by the high performance liquid chromatography and International Federation of Clinical Chemistry reference method. Results The high and positive correlation coefficient was demonstrated between the enzymatic and International Federation of Clinical Chemistry methods for the samples with normal HbF concentration. The samples with high HbF concentration exhibited high and positive correlation coefficient, although the figure is lower than that of the normal HbF concentration group. Conclusions BM Test HbA1c revealed a high and positive correlation with high performance liquid chromatography and International Federation of Clinical Chemistry reference methods. However, the HbA1c values by BM Test HbA1c and high performance liquid chromatography method were lower than those by International Federation of Clinical Chemistry reference method when the HbF concentration was high. BM Test HbA1c had some deviation from the International Federation of Clinical Chemistry reference method, while no significant interference was seen when the HbF values were less than 15%. When HbF is higher than 15%, however, the interpretation of HbA1c by BM Test may require some special consideration.

Publisher

SAGE Publications

Subject

Clinical Biochemistry,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3