Improvement of point of care testing device for accurate whole blood glucose measurement in early neonates

Author:

Koyano Kosuke12ORCID,Arioka Makoto12,Nakao Yasuhiro1,Morimoto Aya1,Sugino Masashiro13ORCID,Morita Hirosuke12,Nakamura Shinji1,Kondo Sonoko1,Konishi Yukihiko1,Yasuda Saneyuki1,Kusaka Takashi1

Affiliation:

1. Department of Pediatrics, Faculty of Medicine, Kagawa University, Kagawa, Japan

2. Maternal and Perinatal Center, Kagawa University Hospital, Kagawa University, Kagawa, Japan

3. Division of Neonatology, NHO Shikoku Medical Center for Children and Adults, Zentsuji City, Japan

Abstract

Background It is important that blood glucose concentrations be accurately and conveniently measured in infants. However, especially in the early neonatal period, point-of-care testing devices used for adults may not accurately measure blood glucose concentrations in neonates. Methods In Study 1, the accuracy of neonatal whole-blood glucose measurements was evaluated for the existing glucose analyser Glutest Mint® (hereinafter MINT1; Sanwa Kagaku Kenkyusho, Nagoya, Japan) by comparing the data with reference blood glucose concentrations. In Study 2, we used MINT2, which was modified based on the findings from Study 1, to measure whole-blood glucose concentrations in newborns, and the accuracy of the measurements was compared with that of MINT1. Results Blood glucose concentrations were measured in 100 infants each in Study 1 and 2. In Study 1, the whole-blood glucose concentrations measured using MINT1 were found to be significantly lower than the reference blood glucose concentrations in early neonates. The results of Study 1 suggested that characteristics of erythrocyte membranes in early neonates affected the measurements. Therefore, we conducted Study 2 using MINT2, which was modified to be less susceptible. MINT2 was found to accurately measure whole-blood glucose concentrations in the early neonatal period. Conclusion The study showed that the point-of-care testing device could be improved to allow for accurate whole-blood glucose measurements in the early neonatal period.

Funder

PHC Holding Corporation

Publisher

SAGE Publications

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Corrigendum to “Improvement of point of care testing device for accurate whole blood glucose measurement in early neonates”;Annals of Clinical Biochemistry: International Journal of Laboratory Medicine;2024-07-31

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3