Immunoturbidimetric Measurement of Serum Retinol-Binding Protein in Renal and Hepatic Disease

Author:

Bankson D D,Rifai N,Silverman L M

Abstract

Serum retinol-binding protein (RBP), with a biological half-life of less than 12 h, is a useful indicator of liver or kidney dysfunction. An automated immunoturbidimetric assay for the measurement of RBP has been developed using the Cobas-BIO centrifugal analyser and commercially available materials. Serum samples with RBP concentrations as low as 3 mg/L were measured. Within-run and day-to-day imprecision were 3·7 and 5·7%, respectively. The reference range (mean±2SD) for 51 adults was 17 to 61 mg/L. Slight haemolysis of serum (haemoglobin 10 g/L) resulted in an apparent 8% reduction of RBP with greater interference at higher haemoglobin concentrations. However, bilirubin in concentrations up to 0·15 g/L did not interfere with RBP measurements. There was good correlation between immunoturbidimetry and a commercial radial immunodiffusion method. Serum RBP concentrations were decreased in liver disease and increased in renal failure.

Publisher

SAGE Publications

Subject

Clinical Biochemistry,General Medicine