Affiliation:
1. Department of Clinical Chemistry, Wolfson Research Laboratories, Queen Elizabeth Medical Centre, Edgbaston, Birmingham B15 2TH
Abstract
The Greiner Selective Analyzer (GSA II) was evaluated over a period of six months. The evaluation assessed the reliability, accuracy, and precision of the analyser for six determinations. The methods evaluated were for glucose, urea, creatinine, total protein, total bilirubin, and alkaline phosphatase. Comparison of results was also made with those obtained for the same specimens using the Technicon SMA 12/60 Analyzer. Correlation and comparison of results indicate that the Greiner Selective Analyzer performed better for three of the methods but worse for serum creatinine determination. The role of the analyser as a routine tool in the clinical laboratory was also evaluated during analyses of approximately 900 patient specimens. Other features evaluated were analytical range of the six methods under study, the economics of operation, temperature control, and electrical and mechanical safety.
Subject
Clinical Biochemistry,General Medicine
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Browsing the seventies;Annals of Clinical Biochemistry: International Journal of Laboratory Medicine;2013-07-25
2. Measurement of Alkaline Phosphatase Activity;Alkaline Phosphatase;1979
3. A Long-Term Evaluation of the Centrifichem Centrifugal Microanalyser during Routine Use in the Clinical Biochemistry Laboratory;Annals of Clinical Biochemistry: International Journal of Laboratory Medicine;1977-01