Affiliation:
1. Department of Chemical Pathology, St James's University Hospital, Leeds LS9 7TF, UK
Abstract
The ionised calcium concentration of sequential retentates prepared by ultrafiltration of a human control serum increased with increasing protein concentration when measured with both a Nova 2 and a Radiometer ICA1 analyser using their standard reference electrodes. In contrast, the ionised calcium in the same retentates fell slightly with increasing protein when the reference electrode liquid junctions of the instruments were changed from hypertonic to isotonic solutions, the values then paralleling those in the filtrates. Thus, the clinically significant positive relationship between ionised calcium and protein that has been reported with the Nova 2 and ICA1 analysers is almost certainly an effect of protein on the reference electrode liquid junction potential rather than a consequence of a Donnan effect on true ionised calcium distribution.
Subject
Clinical Biochemistry,General Medicine
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Diagnostic Tests for Vascular Calcification;Advances in Chronic Kidney Disease;2019-11
2. Ionized calcium measurements are influenced by albumin – should ionized calcium be corrected?;Scandinavian Journal of Clinical and Laboratory Investigation;2014-05-05
3. Calcium adjustment equations in neonates and children;Annals of Clinical Biochemistry: International Journal of Laboratory Medicine;2012-04-25
4. Bone and Mineral Metabolism;Tietz Textbook of Clinical Chemistry and Molecular Diagnostics;2012
5. ISFET and Fiber Optic Sensor Technologies: In Vivo Experience for Critical Care Monitoring;Chemical Reviews;2008-01-08