Affiliation:
1. Department of Clinical Biochemistry, East Birmingham Hospital, Bordesley Green, Birmingham B9 5ST, UK
Abstract
The aim of this study was to find a method for creatinine analysis unaffected by jaundice by comparing four different methods with high performance liquid chromatography (HPLC) as a reference method. The methods investigated were Kodak DTSC single slide, Boehringer alkaline picrate with and without blank correction and a Boehringer enzymatic method, the last three using the Hitachi 737 analyser. HPLC results were compared with the results of the methods studied. In addition, graphs plotting the difference from the HPLC result against the bilirubin concentration were made; the ideal method would yield a slope and intercept of zero. While all methods showed a positive bias compared with HPLC, the Kodak slide and the Boehringer alkaline picrate with blank correction methods eliminated bilirubin interference and we have changed to the latter method as a consequence of this work.
Subject
Clinical Biochemistry,General Medicine
Cited by
17 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献