Affiliation:
1. Beckman Instruments, 92-94 Chemin des Bourdons, 93220, Gagny, France
Abstract
Automated immunonephelometric assays were developed to measure human IgG subclasses in serum, on the Beckman Array Protein System (APSR), with sheep antihuman IgG subclass antisera. The interassay imprecision was judged to be satisfactory in each case (CV: IgG,: 2·3%; IgG2: 2·9%; IgG3: 4·0% and IgG4 5·4%). The standard curves were found to be linear in the ranges 1·5–30·0 g/L for IgG1, 1·5–15·0 g/L for IgG2, 0·1–1·0 g/L for IgG3 and 0·1–1·0 g/L for IgG4. The lower limit for quantification of the immunonephelometric assays was 0·50 g/L for IgG1, 0·04 g/L for IgG2, 0·06 g/L for IgG3 and 0·03 g/L for IgG4. We found no antigen excess to at least 60·0 g/L for IgG, and IgG2, 3·5 g/L for IgG3 and 2·0 g/L for IgG4. Nephelometric results correlated with those of radial immunodiffusion (r = 0·96 for IgG1, r = 0·93 for IgG2, r = 0·90 for IgG3 and r = 0·96 for IgG4). The reaction monitored by the Beckman Array Protein System is easy to perform, rapid and precise. The performance makes the immunonephelometric method suitable for the determination of IgG subclass concentrations in routine analysis.
Subject
Clinical Biochemistry,General Medicine
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献