Affiliation:
1. Department of Chemical Pathology, St. George's Hospital, London SW17
Abstract
Measurement of enzyme activity in a Clinicon reaction rate analyser was carried out in two ways: one method was based on absolute calibration and the other used an enzyme standard which involved adjustment of results according to the value for the standard obtained in each batch. Batch calibration significantly improved long-term precision and reduced differences between results obtained in different instruments.
Subject
Clinical Biochemistry,General Medicine
Cited by
1 articles.
订阅此论文施引文献