Sodium Measurements in the Presence of Paraproteins by Four Direct ISE Methods and Flame Photometry Compared

Author:

Burnett David1,Ayers Gwendolen J1,Rumjen Sherab C1,Woods Terence F1

Affiliation:

1. Department of Clinical Biochemistry, St Albans City Hospital, Normandy Road, St Albans, Herts, UK

Abstract

Sodium was determined by flame photometry and by direct potentiometry using four different systems (Ektachem DT60, Corning 614, Orion 1020 and Nova-1) in serum samples from patients with normal proteins and paraproteins. Flame photometric and direct potentiometric differences (ISE-flame), showed the expected relationship with total protein whether or not the sample contained paraprotein, an exception being the Ektachem DT60 on normal protein. With the exception of the Ektachem DT60 with samples with normal protein levels, correlation between different ISE-flame estimates was good. Differences between direct potentiometric measurements are probably more attributable to efforts made by manufacturers to relate them to flame photometric measurements in ‘normal samples’, than to the volume occupied by increased protein content. The apparent lack of influence of protein on ISE-flame for the Ektachem DT60 was shown to relate to the original external reference electrode salt bridge used. A second study showed that changing this fluid to an equi-transferrant high salt content caesium chloride based reference fluid partially unmasked the protein effect.

Publisher

SAGE Publications

Subject

Clinical Biochemistry,General Medicine

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3