Affiliation:
1. Department of Special Care Baby Unit (SCBU) Arrowe Park Hospital, Wirral L49 5PE, UK
Abstract
Near patient testing for glucose is now a widely accepted procedure in hospital wards and clinics. However, in a neonatal ward where the detection of hypoglycaemia rather than hyperglycaemia is of paramount importance, it is more difficult to find a suitable glucose monitoring instrument. We compared two Hemocue blood glucose analysers (Hemocue Ltd) in our special care baby unit (SCBU) with the laboratory procedures and found that the Hemocue may overestimate the glucose by as much as 2·5mmol/L. In addition, Hemocue analysers are costly to run. We feel these analysers may be more useful in a general ward rather than in a SCBU.
Subject
Clinical Biochemistry,General Medicine
Cited by
17 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献