Background: The objective of this study was to compare the mechanical performance of 4 different constructs for fixation of oblique scaphoid fractures. Methods: Twenty-eight synthetic scaphoids underwent an oblique osteotomy along the dorsal sulcus. Each was randomly assigned to fixation by 1 of 4 methods: two 1.5-mm headless compression screws, one 2.2-mm screw, one 3-mm screw, or a 1.5-mm volar variable-angle plate. After fixation, scaphoids were potted at a 45° angle and loaded at the distal pole by a hydraulically driven mechanical testing system plunger until the fixation failed. Excursion and load were measured with a differential transformer and load cell, respectively. From these data, the stiffness, load-to-failure, and maximum displacement of each construct were calculated. Results: The 2.2-mm screw demonstrated the highest stiffness and the two 1.5-mm screws had the lowest. However, there were no significant differences among the fixation methods in terms of stiffness. Both 2.2- and 3-mm screw constructs had significantly higher loads-to-failure than two 1.5-mm screws. The maximum load for the plate approached, but did not achieve, statistical significance compared with the 1.5-mm screws. There was no significant difference among constructs in displacement. Conclusions: All constructs demonstrated similar mechanical properties that may provide sufficient stability for effective clinical use. Given their significantly higher loads-to-failure, a 2.2- or 3-mm screw may be superior to two 1.5-mm screws for fixation of unstable scaphoid fractures. The volar plate did not have superior mechanical characteristics to the compression screws.
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
15 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献