Determining Measurement Error for Bohler’s Angle and the Effect of X-Ray Obliquity on Accuracy

Author:

Gonzalez Tyler A.1234,Ehrlichman Lauren K.1234,Macaulay Alec A.1234,Gitajn I. Leah1234,Toussaint R. James1234,Zurakowski David1234,Kwon John Y.1234

Affiliation:

1. Department of Orthopaedic Surgery, Massachusetts General Hospital, Boston, Massachusetts (TAG, LKE, AAM, ILG)

2. The Orthopaedic Institute, Gainesville, Florida (RJT)

3. Orthopedic Center, Boston Children’s Hospital, Boston, Massachusetts (DZ)

4. Department of Orthopaedic Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts (JYK)

Abstract

Background. Bohler’s angle (BA) is the most commonly utilized radiographic measurement in the study of calcaneus fractures and has been shown to be prognostic in nature. Therefore, it is critical that the measurement of BA be accurate as both therapeutic and prognostic information relies on it. Oblique lateral radiographs can be a cause of error in BA measurements. However, measurement error and the effects of X-ray beam obliquity on BA have not been established in the literature. The purpose of this study was to determine measurement error and understand the effects of X-ray beam’s obliquity on the measurement of BA. Methods. A cadaver specimen was imaged using a C-arm to obtain a perfect lateral radiograph of the ankle and slightly oblique lateral views in the anterior, posterior, cephalad, and caudad directions in 5° increments (21 images). Metallic beads were then placed on the anterior calcaneal process, posterior facet, and the superior aspect of the posterior tuberosity, and the same 21 images were then obtained. The metallic beads placed on the reference radiographs allowed the authors to accurately measure BA for each image and served as reference for the corresponding test radiographs. Thirty-four orthopaedic staff members participated in the study and used DICOM measurement tool to measure BA on each of the 21 test radiographs. The measurements were then compared to the measurements of BA from the reference radiographs to determine error in measurement. Results. A total of 714 different measurements were obtained. Average measurement error was 6° (95% confidence interval = −4° to 15°). The difference between the observed BA measurements compared to the true BA measurements increased with increasing X-ray obliquity. Conclusions. Measurement error for BA is ±6° and increases most with cephalad oblique radiographs. Orthopaedic surgeons’ ability to accurately measure BA significantly decreases with increasing obliquity of the lateral radiograph. Levels of Evidence: Level V: Cadaver bench study

Publisher

SAGE Publications

Subject

Orthopedics and Sports Medicine,Podiatry,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3