Affiliation:
1. Department of Mechanical Engineering, Fujian Agriculture and Forestry University, Fuzhou, China
Abstract
This paper proposes a novel capsule robot (CR) with spring-connected legs for potential usage in the gastrointestinal (GI) tract. The proposed CR has a set of six legs and is actuated by a micromotor through internal mechanisms. Every two adjacent legs are connected by a spring. The leg opening-folding movement, together with the spring stretching-contraction motion, may be employed to distend the GI tract, carry out drug delivery or perform biopsy. The relationship between the driving force required by the spring set and the spring length was carefully calculated and simulated. This relationship was also measured by an NI data acquisition system. The calculated and measured data showed good accordance with each other. In addition, the driving force from the micromotor was calculated. Both the calculations and the CR experiments demonstrated that the internal motor could provide sufficient force on the springs.
Funder
Natural Science Foundation of Fujian Province
Scientific Research Fund of Fujian Provincial Education Department