Unstable Vancouver B1 periprosthetic femoral fracture fixation: A biomechanical comparison between a novel C-shaped memory alloy implant and cerclage wiring

Author:

Oh Seog-Hyun1ORCID,Suh Yu-Sung2,Eghan-Acquah Emmanuel1,Yurevich Kollerov Mikhail3,Won Sung-Hun2,Baek Min Jung4,Lee Sung-Jae1,Choi Sung-Woo2ORCID

Affiliation:

1. Department of Biomedical Engineering, Inje University, Gimhae, 50834, 197 Inje-ro, Gimhae-si, Gyeongsangnam-do, Republic of Korea

2. Department of Orthopedic Surgery, Soonchunhyang University College of Medicine, Seoul, 04401, 59 Daesagwan-ro, Yongsan-gu, Seoul, Republic of Korea

3. Department of Materials Science and Materials Technology, Moscow Aviation Institute, 121552, Kom.1, Dom 5, Orshanskaya Str. Moscow, Russia

4. Department of Obstetrics and Gynecology, Bundang CHA Hospital, Seongnam, 13496, 59 Yatap-ro, Bundang-gu, Seongnam-si, Gyeonggi-do, Republic of Korea

Abstract

Background To compare the biomechanical stability of a novel, C-shaped nickel–titanium shape memory alloy (SMA) implant (C-clip) with traditional cerclage wiring in the fixation of a Vancouver B1 (VB1) periprosthetic femoral fracture (PFF). Methods In total, 18 synthetic femoral fracture models were constructed to obtain unstable VB1 fracture with an oblique fracture line 8 cm below the lesser trochanter. For each model, the distal portion was repaired using a 10-hole locking plate and four distal bi-cortical screws. The proximal portion was repaired using either three, threaded cerclage wirings or three, novel C-shaped implants. Specimens underwent biomechanical testing using axial compression, torsional and four-point bending tests. Each test was performed on three specimens. Results The C-clip was statistically significantly stronger (i.e., stiffer) than cerclage wiring in the three biomechanical tests. For axial compression, medians (ranges) were 39 (39–41) and 35 (35–35) N/mm, for the C-clip and cerclage wiring, respectively. For torsion, medians (ranges) were, 0.44 (0.44–0.45) and 0.30 (0.30–0.33) N/mm for the C-clip and cerclage wiring, respectively. For the four-point bending test, medians (ranges) were 39 (39–41) and 28 (28–31) N/mm; for the C-clip and cerclage wiring, respectively. Conclusion Results from this small study show that the novel, C-shaped SMA appears to be biomechanically superior to traditional cerclage wiring in terms of stiffness, axial compression, torsion and four-point bending, and may be a valuable alternative in the repair of VB1 PFF. Further research is necessary to support these results.