Biomechanical Comparison of Short-, Mid-, and Long-Length Proximal Femoral Nails for Femoral Intertrochanteric Fracture (AO/OTA 31A3.3) Fixation

Author:

Ando Jiro1ORCID,Takahashi Tsuneari2ORCID,Matsumura Tomohiro3,Takeshita Katsushi1

Affiliation:

1. Department of Orthopedics, School of Medicine, Jichi Medical University, Shimotsuke, Japan

2. Department of Orthopedic Surgery, Ishibashi General Hospital, Shimotsuke, Japan

3. Department of Emergency and Critical Care Medicine, Jichi Medical University, Shimotsuke, Japan

Abstract

Introduction In surgeries for unstable AO/OTA 31A3.3 fractures, surgeons use various lengths of intramedullary nails (IMNs). However, there is insufficient evidence regarding the appropriate nail length for these fractures. This study compared the biomechanical properties of IMNs of different lengths for AO/OTA 31A3.3 fractures. Methods 30 synthetic femora of AO/OTA 31A3.3 fracture model were randomly assigned to the following three groups: short- (170 mm), mid- (235 mm), and long-length (300 mm) nail groups, and were performed fixation surgery. The translation patterns of the constructs were examined by cyclic testing and compared among three groups. Additionally, changes in the neck–shaft and shaft–nail angles after cyclic testing were evaluated using radiological images. Results The translation patterns during cyclic loading did not differ among the groups. Conversely, one-way analysis of variance (ANOVA) revealed a significant difference in the neck–shaft angle change (5.8° ± 1.8°, 2.8° ± 1.3°, and 1.9° ± .9° in the short-, mid-, and long-length groups, respectively; P < .001), and post-hoc analysis revealed that the change was greater in the short-length group than in the mid- and long-length groups ( P < .001 and P < .001, respectively). Furthermore, one-way ANOVA revealed a significant difference in the shaft–nail angle change (3.1° ± 2.1°, 1.4° ± 1.4°, and .1° ± .6° in the short-, mid-, and long-length groups, respectively; P < .001), and post-hoc analysis revealed that the change was greater in the short-length group than in the mid- and long-length groups ( P = .044 and P < .001, respectively). Conclusions Short-length nails were associated with relevant changes in the neck–shaft and shaft–nail angles in our AO/OTA 31A3.3 fracture model. Thus, the selection of mid- or long-length nails instead of short-length nails might be better in IMN surgery for these fractures to prevent postoperative varus deformity.

Publisher

SAGE Publications

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3