Comparison of forehead and finger oximetry sensors during the six minute walk test

Author:

Robertson Lucy1ORCID,Lowry Rachel1,Sylvester Karl1,Parfrey Helen1,Moseley Beverley1,Sheares Karen1,Oates Katrina1

Affiliation:

1. Royal Papworth Hospital, Cambridge, UK

Abstract

Background Measurement of oxygen saturation (SpO2) during the 6 minute walk test (6MWT) could be impacted by the measurement site. Aims To compare SpO2 and heart rate (HR) between forehead and finger sensors during the 6MWT. Sensor readings were also to be compared for signal quality and with capillary blood gas (CBG) pre and post 6MWT. Method 80 subjects with pulmonary vascular disease (PVD) and/or interstitial lung disease (ILD) performed the 6MWT. Pulse oximetry was recorded at 30 s intervals. CBG was taken pre and post 6MWT to determine capillary oxygen saturation (SCO2). Results The forehead sensor recorded higher values for SpO2 ( p < 0.001) and HR ( p < 0.01) compared with the finger sensor during the 6MWT. For both sensors, the demonstrated bias compared to CBG post 6MWT was higher and more variable in subjects who desaturated. During the 6MWT there was a higher occurrence ( p < 0.001) of poor signal quality in the finger sensor compared with the forehead sensor. Conclusion This study suggests that the sensor site can impact pulse oximetry readings. The variance in bias suggests pulse oximetry may not accurately reflect SCO2 measurements particularly in subjects who desaturate during 6MWT.

Publisher

SAGE Publications

Subject

Pulmonary and Respiratory Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3