Clinical monitoring of activated clotting time during cardiothoracic surgery: comparing the Hemochron® Response and Hemochron® Signature Elite

Author:

Hoffmann Roland F1,Horsten Sandra1ORCID,Mariani Massimo A2,de Vries Adrianus J3

Affiliation:

1. Department of Cardiothoracic Surgery, Section Extracorporeal Circulation, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands

2. Department of Cardiothoracic Surgery, Section Thoracic Surgery, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands

3. Department of Anesthesiology, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands

Abstract

Introduction: The Activated Clotting Time (ACT) is commonly used to manage anticoagulation during cardiac surgery. The aim of this study was to compare the older manually operated Hemochron® Response and the automated Hemochron® Signature Elite. Methods: In this observational study the clinically relevant differences of both devices were investigated simultaneously, using duplicate measurements, in 29 patients who underwent a Coronary Artery Bypass Grafting (CABG) or Aortic Valve Replacement (AVR) in order to determine reliability, bias, and to detect which method has the lowest variation. Blood samples were obtained from the arterial line prior to surgery, after administration of 300 IU/kg heparin, 5 minutes after initiation of cardiopulmonary bypass and successively every 30 minutes, and after protamine administration. Results: A total of 202 measurements were performed. Of these 10 measurements were out of range in the Response and 9 in the Elite. About 27 single unstable magnet errors were seen in the Response versus no measurement errors in the Elite. No statistically significant differences between the Response (p = 0.22, Wilcoxon rank) and Elite (p = 0.064) duplicates were observed. The Response values were consistently higher during heparinization than the Elite measurements (p = 0.002, repeated measurements) with an average positive bias of around 56 seconds during heparinization (Bland-Altman). Overall, the coefficient of variation (CoV) increased during heparinization. Conclusion: The Elite was more reliable, but the variation was higher for the Elite than the Response. The observed positive bias in the Response compared to the Elite could affect heparin administration during surgery making the two systems not interchangeable.

Publisher

SAGE Publications

Subject

Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3