Abstract

Introduction: Vacuum assisted venous drainage (VAVD) requires the sealing of the hard-shell venous reservoir, thereby creating circumstances where reservoir pressurization may occur. Manufacturers utilize integrated pressure relief valves (IPRV) to mitigate pressurization risk; however, accidents have been reported even with these devices. We have undertaken a performance evaluation of IPRV’s in a large number of hard-shell venous reservoirs. Methods: Reservoirs were sealed and gas insufflated while measuring reservoir internal pressure. Linear regression models were developed to depict the association between internal pressure and gas inflow rate. External secondary one-way valves (ESOV) were assessed for pressure mitigation performance. An assisted venous drainage survey was circulated to Canadian Clinical Perfusionists. Results: The reservoirs tested were adult ( n = 9, 64%) and pediatric ( n = 5, 36%) designs. Significant variability ( p < 0.001) in internal reservoir pressures (range: 0.04–161.41 mmHg) was observed across the titrated gas inflow rate (0.5–10.0 l/min). The regression models demonstrate excellent predictive performance (SE: 0.008–0.309). ESOV’s reduce the reservoir pressure below that of the IPRV; however, they cannot eliminate reservoir pressurization. The survey showed a majority (91%) of respondents use VAVD, and reservoir pressurization events occur regularly (18%). Conclusions: Significant variability among reservoir’s IPRV to mitigate reservoir pressurization exists. The predictive models are extremely accurate at estimating the internal pressure. ESOV performance limitations moderate their utility as a backup pressure mitigation technique. A significant number of reservoir pressurization events are occurring with the use of VAVD. As a result, standardized communication from manufacturers on the purpose and performance of IPRV is recommended in order to delineate the limitations of these devices.

Publisher

SAGE Publications

Subject

Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3