Human albumin in extracorporeal prime: effect on platelet function and bleeding

Author:

Kamra C1,Beney A2

Affiliation:

1. Memorial University of Newfoundland, St. John’s, NL, Canada

2. Eastern Health, St. John’s, NL, Canada

Abstract

Purpose: Synthetic starches have been positioned as an equivalent substitute for human albumin in extracorporeal prime, with both providing osmotic and oncotic pressure. Another effect of albumin is its ability to coat the synthetic surfaces of an extracorporeal circuit with a biopassivating protein monolayer. Whether this protein biopassivation has any benefit to the patient, assessed by platelet count, platelet function and 24-hour bleeding rate, is considered. Methods: Patients presenting for coronary artery bypass at a Canadian tertiary care hospital were randomized into two groups until a final study size of 20 patients was obtained. The Study Group received 2.5 g of human albumin in the extracorporeal prime and the Control Group remained protein free. Both groups included Voluven 6% as a synthetic starch. Blood samples were obtained at three intervals; Pre-bypass, During bypass (30 minutes after initiation of bypass), and Post-bypass. These samples were assayed for platelet function, platelet count and hemoglobin. Chest tube drainage over a 24-hour period was monitored. Results: Platelet count was significantly higher in the During sample in the Study Group (196 ± 56.5 x 109/ml versus 160 ± 18.5 x 109/ml, p<0.05), however, this difference was no longer significant with the Post-bypass sample (135 ±36.0 x 109/mL versus 127 ±19 x 109/mL). Platelet function assays (PFA) showed no significant differences. Chest tube drainage after 24 hours was significantly lower in the Study Group (586 ± 131.8 ml/24 h versus 741 ± 272.5 ml/24 h, p<0.05). Conclusions: Human albumin can passivate the synthetic surfaces of the extracorporeal circuit, which is supported by observations of preserved platelet count and reduced chest tube drainage. Although some statistically significant benefits were observed, the practical benefits of passivating an extracorporeal circuit with human albumin may be minimal.

Publisher

SAGE Publications

Subject

Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine

Cited by 16 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3