Affiliation:
1. The University Clinic of Internal Medicine III, Cardiology, Innsbruck Medical University, Innsbruck, Austria
Abstract
Post-cardiac catheterization femoral artery hemostasis can be accomplished with several mechanisms, including the FemoSeal® hemostasis device which has been designed and approved for closure of 6 French (F) arterial puncture sites. The aim of this study was to investigate whether the FemoSeal® vascular closure device can effectively and safely seal 7F arterial puncture sites after diagnostic and interventional cardiac catheterizations. Femoral artery puncture sites of 50 consecutive patients undergoing cardiac catheterization were closed with the FemoSeal® vascular closure device, according to the manufacturer’s instructions. Efficacy endpoints were time to hemostasis and successful ambulation. Safety endpoints included bleeding complications, vessel occlusion and pseudoaneurysms. Mean time to hemostasis was 57.8±26.3 seconds (0-125 seconds). Hemostasis was achieved in 100 percent of the 50 patients. One patient suffered minor bleeding the next day, i.e. local hematoma. This clinical study demonstrates that the FemoSeal® vascular closure device, initially approved for closure of 6F arterial puncture sites, shows promising efficacy and safety to seal a larger (7F) femoral arterial puncture sites after diagnostic and interventional cardiac catheterizations.
Subject
Advanced and Specialised Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology Nuclear Medicine and imaging,General Medicine
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献