Optimized performance of the Abiomed BVS 5000: adjustment of the pump height based on Doppler control of the flow pattern

Author:

Lachat M1,Jaggy C2,Leskosek B2,Jenni R3,Büchi M3,Zünd G2,Künzli A2,Vogt P2,Turina M2

Affiliation:

1. Clinic for Cardiovascular Surgery, University Hospital, Zurich,

2. Clinic for Cardiovascular Surgery, University Hospital, Zurich

3. Clinic for Cardiology, University Hospital, Zurich

Abstract

The Abiomed BVS 5000 is an automatic volume-driven paracorporeal pulsatile assist device providing left, right or biventricular support. The paracorporeal position allows optical adjustment of filling volumes of the device, which determines the output of the system. A procedure to adjust for maximal stable flow has not yet been established. In vitro measurements have been performed to assess the flow and pressure characteristics of the Abiomed BVS 5000 by raising the preload in 5 mmHg steps before running the system. Doppler probes were placed at the inflow and outflow lines of the pump. After setting the afterload at 80 mmHg the assist device was started. Two measurements were performed to find optimal flow (based on Doppler control and optical adjustment). (1) By Doppler control a stable flow pattern was found at a preload of 25 mmHg with a mean atrial pressure of 5 mmHg and a mean flow of 5.3 ± 0.7 l/min (mean ± standard deviation) at the inflow and outflow sites (the console flow was 4.8 ± 0.4 l/min with a frequency of 61.8 ± 2.0 l/min). (2) Optical adjustment of the pump height gave rise to a preload of 35 mmHg where we recorded a maximal atrial pressure of 107 ± 5.8 mmHg, a maximal retrograde flow of -4.3 ± 1.2 l/min at the inflow and -1.2 ± 0.4 l/min at the outflow site. The mean flow at the inflow and outflow sites was 5.1 ± 0.5 l/min (the console flow was 4.6 ± 0.3 l/min with a frequency of 59.6 ± 2.6 Hz). At an initial afterload of 60 and 40 mmHg the system showed the same qualitative behaviour, but the results were less accurate. Optical adjustment of the pump height may result in an atrioventricular valve insufficiency with undetected retrograde flow and high atrial pressures. We conclude that a Doppler flow probe must be placed at the inflow site to guarantee maximal stable flow.

Publisher

SAGE Publications

Subject

Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3