Adding an extension piece to the end of the purge side arm of the Impella device can prevent the incidence of the cassette breaking and decrease the Impella device failure rate: Impact of practice change on patient outcome

Author:

Satashia Parthkumar1,White Andrew1,Isha Shahin1,Hanson Abby1ORCID,Jenkins Anna1ORCID,Blasavage Jessica1,Matos Nikki1,Tomlinson Amanda1,Zhang Stephanie1,Quinones Quintin1,Waldron Nathan1,Bhattacharyya Anirban1,Kiley Sean1,Guru Pramod1ORCID,Chaudhury Sanjay1,Shapiro Anna1,Moreno Franco Pablo1,Sanghavi Devang K1

Affiliation:

1. Department of Critical Care Medicine, Mayo Clinic in Florida, Jacksonville, FL, USA

Abstract

Background: Impella 5.5® with Smart Assist is a minimally invasive Left Ventricular Assist Devices (LVAD) approved by the Food and Drug Administration (FDA) for treating ongoing cardiogenic shock for up to 14 days. The Impella® intends to reduce ventricular workload and provide the circulatory support necessary for myocardial recovery. Research Question: Compared to standard practice, does adding an extension piece to the purge tube side arm of the Impella® Device decrease the incidence of device failure and positively impact the health outcome of adult patients receiving Impella® support? Study Design and Methods: A retrospective chart review of ICU patients was done at a tertiary care center from August 2018 to August 2022 to assess the differences in patient outcomes related to Impella® Device utilization before and after the implementation of the extension piece to the purge tube sidearm. Among patients reviewed, a total of 20 were included in our review, with seven not having the purge tube side arm extension added, while 13 patients had the extension. Results: The two study groups had no significant difference in patient health outcomes. Additionally, there were no instances of device failure requiring explanation without the extension tubing. However, there were no cases of the purge cassette cracking with the addition of the extension tubing. Conclusion: The addition of extension tubing to the purge cassette of the Impella® Device did not impact patient health outcomes or the incidence of device failure. There was a complete reduction in the incidence of the purge cassette cracking, which could reduce the potential for infection or device failure over a long period of mechanical support. There is a need for long-term prospective studies to confirm the results.

Publisher

SAGE Publications

Subject

Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3