Affiliation:
1. Department of Cardiac Surgery, Royal Hospital for Sick Children, Yorkhill, Glasgow
Abstract
The clinical measurement of hepatic perfusion is complicated by a dual blood supply and the invasive nature of the majority of techniques available. The aim of this study was to validate indocyanine green clearance and noninvasive auricular densitometry as a measure of hepatic perfusion in the context of paediatric cardiac surgery. The effects of different dye concentrations on densitometer recording were assessed in vitro and found to have a linear relationship. Similarly, variations in haematocrit, within the range 21-47%, also had little effect on accuracy. Comparison of densitometry and direct blood sampling with plasma spectrophotometry in six postoperative, normothermic children showed no significant difference between the noninvasive and invasive techniques (r = 0.968; p > 0.05, t-test). Comparison in 10 hypothermic children during cardiopulmonary bypass also showed no significant difference between the two methods, provided that no further cooling or rewarming took place (r = 0.83; p > 0.05, Wilcoxon test). Noninvasive auricular densitometry can, therefore, provide a reliable assessment of hepatic perfusion in children undergoing cardiac surgery.
Subject
Advanced and Specialised Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology Nuclear Medicine and imaging,General Medicine
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献