Affiliation:
1. Department of Paediatric Surgery, Ostra Hospital, University of Göteborg, Göteborg
2. Department of Medicine, Sahlgrenska University Hospital, University of Göteborg, Göteborg
3. Department of Paediatric Surgery, Ostra Hospital, University of Göteborg Göteborg
Abstract
Blood platelets are rapidly activated in contact with biomaterials and, therefore, can be used as markers of the biocompatibility of various components in an extracorporeal system. In the present work, two different oxygenators, one membrane oxygenator (Avecor) and one hollow-fibre oxygenator ('Lilliput', Dideco) were compared. Complete in vitro extracorporeal membrane oxygenation circuits were perfused with fresh, heparinized human blood for 24 h. Eight experiments were performed with the hollow-fibre oxygenator and five experiments with the membrane oxygenator. Blood gases, electrolytes, glucose and haematocrit were kept within physiological limits. Platelet count, plasma concentration of β-thromboglobulin, platelet serotonin content, platelet membrane glycoprotein Ib and its degradation product glycocalicin, as well as plasma haemoglobin concentration were assayed. As regards most of these variables, significant differences in favour of the hollow- fibre oxygenator were observed.
Subject
Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献