Hearing Aid Self-Adjustment: Effects of Formal Speech-Perception Test and Noise

Author:

Mackersie Carol L.1ORCID,Boothroyd Arthur12ORCID,Garudadri Harinath2

Affiliation:

1. School of Speech, Language and Hearing Sciences, San Diego State University

2. Qualcomm Institute of Calit2, University of California, San Diego

Abstract

While listening to recorded sentences with a sound-field level of 65 dB SPL, 24 adults with hearing-aid experience used the “Goldilocks” explore-and-select procedure to adjust level and spectrum of amplified speech to preference. All participants started adjustment from the same generic response. Amplification was provided by a custom-built Master Hearing Aid with online processing of microphone input. Primary goals were to assess the effects of including a formal speech-perception test between repeated self-adjustments and of adding multitalker babble (signal-to-noise ratio +6 dB) during self-adjustment. The speech test did not affect group-mean self-adjusted output, which was close to the National Acoustics Laboratories’ prescription for Non-Linear hearing aids. Individuals, however, showed a wide range of deviations from this prescription. Extreme deviations at the first self-adjustment fell by a small but significant amount at the second. The multitalker babble had negligible effect on group-mean self-selected output but did have predictable effects on word recognition in sentences and on participants’ opinion regarding the most important subjective criterion guiding self-adjustment. Phoneme recognition in monosyllabic words was better with the generic starting response than without amplification and improved further after self-adjustment. The findings continue to support the efficacy of hearing aid self-fitting, at least for level and spectrum. They do not support the need for inclusion of a formal speech-perception test, but they do support the value of completing more than one self-adjustment. Group-mean data did not indicate a need for threshold-based prescription as a starting point for self-adjustment.

Publisher

SAGE Publications

Subject

Speech and Hearing,Otorhinolaryngology

Reference41 articles.

1. Direct-to-Consumer Hearing Devices: Capabilities, Costs, and Cosmetics

2. ANSI (2014).Specification of hearing aid characteristics (ANSI S3.22-2014) New York, NY, American National Standards Institute.

3. ANSI (1997).Methods for the calculation the speech intelligibility index (ANSI S3.5 -1997 R2012) New York, NY, American National Standards Institute.