Image-quality characteristics in the longitudinal direction from different image-reconstruction algorithms during single-rotation volume acquisition on head computed tomography: A phantom study

Author:

Watanabe Ryo1ORCID,Zensho Ayako1,Ohishi Yoshitaka1,Funama Yoshinori2ORCID

Affiliation:

1. Department of Radiology, Hospital of the University of Occupational and Environmental Health, Kitakyushu, Japan

2. Department of Medical Radiation Sciences, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan

Abstract

Background A multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation. Purpose To investigate variations in image-quality characteristics in the longitudinal direction for different image-reconstruction algorithms and strengths with phantoms. Material and methods Single-rotation volume scans were performed on a 320-row multidetector CT volume scanner using three types of phantoms. Tube current was set to 200 mA (standard dose) and 50 mA (low dose). All images were reconstructed with filtered back projection (FBP), mild and strong levels with hybrid iterative reconstruction (HIR), and model-based IR (MBIR). Computed tomography numbers, image noise, noise power spectrum (NPS), task-based transfer function (TTF), and visual spatial resolution were used to evaluate uniformity of image quality in the longitudinal direction ( Z-axis). Results The MBIR images showed smaller variation in CT numbers in the Z-axis. The difference in the highest and lowest CT numbers was smaller (5.0 Hounsfield units [HU]) for MBIR than for FBP (6.6 HU) and HIR (6.8 HU). The variations in image noise were the smallest for strong MBIR and the largest for FBP. The low-frequency component at NPS0.2 was lower for strong MBIR than for other algorithms. The high-frequency component at NPS0.8 was low in all reconstructions. For MBIR, the image resolution and TTFs were higher in the outer portion than in the center. Conclusion Model-based IR is the optimal image-reconstruction algorithm for single-volume scan of spherical subjects owing to its high in-plane resolution and uniformity of CT numbers, image noise, and NPS in the Z-axis.

Publisher

SAGE Publications

Subject

General Earth and Planetary Sciences,General Environmental Science

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3