Teflon-Wire Piston or Stainless-Steel Bucket Stapes Prosthesis: Does it Make a Difference?

Author:

Farrior Jay B.1,Temple Anne E.2

Affiliation:

1. Farrior Ear Clinic, Tampa, Fla.; Department of Otolaryngology, University of Florida, Tampa.

2. Farrior Ear Clinic, Tampa, Florida

Abstract

The goal of this study was to determine whether postoperative (implantation of a stapes prosthesis) hearing gain and the amount of air-bone gap overclosure are more improved with the Teflon-wire piston or with the stainless-steel bucket prosthesis. We retrospectively reviewed the outcomes of 82 surgeries that had been performed by the primary author; 41 of these patients had received a Fisch Teflon-wire piston, and 41 had received a Bailey-modified Robinson stainless-steel bucket prosthesis. The mean hearing gain for the patients who received the Teflon-wire piston was 23.3 dB after primary stapes surgery and 20.5 dB after revision surgery. Patients who received the stainless-steel bucket prosthesis experienced a mean hearing gain of 20.7 and 20.3 dB, respectively. Following primary stapes surgery, the air-bone gap overclosure was 4.4 dB with the Teflon-wire piston and 5.2 dB with the stainless-steel bucket prosthesis. There was no statistically significant difference in either hearing gain or air-bone gap overclosure between the two prostheses.

Publisher

SAGE Publications

Subject

Otorhinolaryngology

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Stapes surgery: a National Survey of British Otologists;European Archives of Oto-Rhino-Laryngology;2015-02-25

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3