Affiliation:
1. Permedica S.p.A., Merate, Italy
2. Endolab Mechanical Engineering GmbH, Rosenheim, Germany
Abstract
Background Glenoid component failures still represent the most common complication in total shoulder arthroplasty. These failures depend on several factors, including ultra-high molecular weight polyethylene (UHMWPE) wear. One reason for UHMWPE wear in total shoulder arthroplasty may be the current use of a spherical prosthetic humeral head against a radially mismatched UHMWPE glenoid component, which leads to reduced glenohumeral translations, glenoid edge loading and high translational forces during shoulder motions. The aim of this study was to assess the in vitro wear of an anatomic total shoulder prosthesis with non-spherical non-conforming bearings with inverted conventional materials. Methods The wear of a vitamin E-blended UHMWPE non-spherical humeral head articulating against a non-conforming titanium-niobium nitride (TiNbN)-coated metallic glenoid was tested using a joint simulator. The wear test was performed by applying a constant load of 756 N with angular motions and translations. Results After 2.5 million cycles, the mean wear rate of the humeral head was 0.28 ± standard deviation (SD) 0.45 mg/million cycles. Conclusion The low wear rate of the vitamin E UHMWPE humeral head supports the use of non-spherical non-conforming bearings with inverted conventional materials in anatomic total shoulder arthroplasty.
Subject
Rehabilitation,Physical Therapy, Sports Therapy and Rehabilitation,Orthopedics and Sports Medicine,Surgery
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献